lexiara

7.2.

The instruments shall meet the metrological requirements within the temperature range specified by the manufacturer. The value of this range shall be at least equal to: 5 °C for an instrument in class I; 15 °C for an instrument in class II; 30 °C for an instrument in class III or IIII. In the absence of a manufacturer’s specification, the temperature range of – 10 °C to + 40 °C applies.