7.2.2.

The error of indication for the volume metered after the durability test shall not exceed: ± 6 % of the metered volume between Q_(1) included and Q_(2) excluded; ± 2,5 % of the metered volume between Q_(2) included and Q_(4) included for water meters intended to meter water with a temperature between 0,1 °C and 30 °C, ± 3,5 % of the metered volume between Q_(2) included and Q_(4) included for water meters intended to meter water with a temperature between 30 °C and 90 °C.