4.1. Class 1,5 $3

The variation of the measurement result after the durability test when compared with the initial measurement result for the flow rates in the range Q_(t) to Q_(max)- shall not exceed the measurement result by more than 2 %. The error of indication after the durability test shall not exceed twice the MPE in point 2.