4.4. Verification scale interval — multi-interval instruments
Table 4 For i = r the corresponding column of Table 3 applies with e replaced by e_(r). Accuracy classes Verification scale interval Number of verification scale intervals nMaxe Minimum value nMax_(i)e_((i 1)) Maximum value nMax_(i)e_(i) XI Y(I) 0,001 g ≤ e_(i) 50000 — XII Y(II) 0,001 g ≤ e_(i) ≤ 0,05 g 5000 100000 0,1 g ≤ e_(i) 5000 100000 XIII Y(a) 0,1 g ≤ e_(i) 500 10000 XIIII Y(b) 5 g ≤ e_(i) 50 1000 Where: i= 1, 2, … r i= partial weighing range r= total number of partial ranges