2.3. Error relative to pre-set value (setting error)
For instruments where it is possible to pre-set a fill weight; the maximum difference between the pre-set value and the average mass of the fills shall not exceed 0,312 of the maximum permissible deviation of each fill from the average, as specified in Table 5.
← anx_VIII__cpt_3__cpt_3__cpt_2__cpt_2__text_1 · All articles · anx_VIII__cpt_3__cpt_3__cpt_2__cpt_3__text_1 →
Source: EUR-Lex (Cellar) · retrieved 2026-11-06 · Text as adopted (Official Journal); later amendments are not incorporated in this text.