8. Performance under influence factors and electromagnetic disturbances

The MPEs due to influence factors shall be as specified in Table 7. Table 7 Load (m) in totalisation scale intervals (d_(t)) MPE 0 < m ≤ 500 ± 0,5 d_(t) 500 < m ≤ 2000 ± 1,0 d_(t) 2000 < m ≤ 10000 ± 1,5 d_(t) The critical change value due to a disturbance is one totalisation scale interval for any weight indication and any stored total.