5. Performance under influence factor and electromagnetic disturbance

The MPE due to an influence factor shall be as specified in Table 11. Table 11 Load (m) in verification scale intervals (d) MPE 0 < m ≤ 500 ± 0,5 d 500 < m ≤ 2000 ± 1,0 d 2000 < m ≤ 10000 ± 1,5 d The critical change value due to a disturbance is one scale interval.