(b) Associated measuring instruments

Associated measuring instruments shall have an accuracy at least as good as the values in Table 4: Table 4 MPE on Measurements Accuracy classes of the measuring system 0,3 0,5 1,0 1,5 2,5 Temperature ± 0,3 °C ± 0,5 °C ± 1,0 °C Pressure Less than 1 MPa: ± 50 kPa From 1 to 4 MPa: ± 5 % Over 4 MPa: ± 200 kPa Density ± 1 kg/m^(3) ± 2 kg/m^(3) ± 5 kg/m^(3) These values apply to the indication of the characteristic quantities of the liquid displayed by the conversion device.