CHAPTER III
Dimensions within the range specified by the manufacturer for the instrument. The manufacturer shall specify the limitations of the instruments due to the speed, and thickness of the surface conditions if relevant, of the product. The MPE is 1,0 %, but not less than 1 dm^(2). In the case of pulling back or stopping the product, it shall not be possible to have an error of measurement or the display must be blanked. The instruments must have a scale interval of 1,0 dm^(2). In addition, it must be possible to have a scale interval of 0,1 dm^(2) for testing purposes.