1.5.4. Calibration
Each normally used operating range shall be calibrated. Using purified synthetic air (or nitrogen), the CO, CO2, NOx and HC analysers shall be set at zero. The appropriate calibration gases shall be introduced to the analysers, the values recorded, and the calibration curve established according to Section 1.5.5. The zero setting shall be rechecked and the calibration procedure repeated, if necessary.