1.5.5.2. Calibration below 15 % of full scale
The analyser calibration curve shall be established by at least 4 additional calibration points (excluding zero) spaced nominally equally below 15 % of full scale. The calibration curve is calculated by the method of least squares. The calibration curve must not differ by more than ± 4 % from the nominal value of each calibration point and by more than ± 1 % of full scale at zero.