1.6. Verification of the calibration

Each normally used operating range shall be checked prior to each analysis in accordance with the following procedure. The calibration shall be checked by using a zero gas and a span gas whose nominal value is more than 80 % of full scale of the measuring range. If, for the two points considered, the value found does not differ by more than ± 4 % of full scale from the declared reference value, the adjustment parameters may be modified. Should this not be the case, a new calibration curve shall be established in accordance with Section 1.5.5.