1.8.2. Hydrocarbon response factors

The analyser shall be calibrated using propane in air and purified synthetic air, according to Section 1.5. Response factors shall be determined when introducing an analyser into service and after major service intervals. The response factor (Rf) for a particular hydrocarbon species is the ratio of the FID C1 reading to the gas concentration in the cylinder expressed by ppm C1. The concentration of the test gas must be at a level to give a response of approximately 80 % of full scale. The concentration must be known to an accuracy of ± 2 % in reference to a gravimetric standard expressed in volume. In addition, the gas cylinder must be preconditioned for 24 hours at a temperature of 298 K ± 5 K (25 °C ± 5 °C). The test gases to be used and the recommended relative response factor ranges are as follows: methane and purified synthetic air 1,00 ≤ Rf ≤ 1,15 propylene and purified synthetic air 0,90 ≤ Rf ≤ 1,10 toluene and purified synthetic air 0,90 ≤ Rf ≤ 1,10 These values are relative to the response factor (Rf) of 1,00 for propane and purified synthetic air.