sec_1_8_4_2__para
Ethane calibration gas shall be flown through the FID with and without bypassing the NMC and the two concentrations recorded. The efficiency shall be determined as follows
← 1.8.4.2. Ethane efficiency · All articles · sec_1_8_4_2__para__2 →
Source: EUR-Lex (Cellar) · retrieved 2026-11-01 · Text as adopted (Official Journal); later amendments are not incorporated in this text.