2.3.1. Data analysis
The air flowrate (Qs) at each restriction setting (minimum eight settings) shall be calculated in standard m3/min from the flowmeter data using the manufacturer's prescribed method. The calibration coefficient shall be calculated from the calibration data for each setting as follows: where, To determine the range of critical flow, Kv shall be plotted as a function of venturi inlet pressure. For critical (choked) flow, Kv will have a relatively constant value. As pressure decreases (vacuum increases), the venturi becomes unchoked and Kv decreases, which indicates that the CFV is operated outside the permissible range. For a minimum of eight points in the region of critical flow, the average Kv and the standard deviation shall be calculated. The standard deviation shall not exceed ± 0,3 % of the average KV.