3.2. Checking of the opacimeter
The zero and full scale checks shall be made in the opacity readout mode, since the opacity scale offers two truly definable calibration points, namely 0 % opacity and 100 % opacity. The light absorption coefficient is then correctly calculated based upon the measured opacity and the LA, as submitted by the opacimeter manufacturer, when the instrument is returned to the k readout mode for testing. With no blockage of the opacimeter light beam, the readout shall be adjusted to 0,0 % ± 1,0 % opacity. With the light being prevented from reaching the receiver, the readout shall be adjusted to 100,0 % ± 1,0 % opacity.