3.2. Flow measurement

The calibration of gas flow meters or flow measurement instrumentation shall be traceable to international and/or national standards. The maximum error of the measured value shall be within ± 2 % of reading. If the gas flow is determined by differential flow measurement, the maximum error of the difference shall be such that the accuracy of GEDF is within ± 4 % (see also Annex V, Section 2.2.1, EGA). It can be calculated by taking the Root-Mean-Square of the errors of each instrument.