sec_1_6__para__3
If, for the two points considered, the value found does not differ by more than ± 4 % of full scale from the declared reference value, the adjustment parameters may be modified. Should this not be the case, a new calibration curve shall be established in accordance with Section 1.5.5.
← sec_1_6__para__2 · All articles · 1.7. Efficiency test of the NOx converter →